Corresponding author
Nobuko Utsumi, Department of Radiation Oncology, Saitama Medical Center, Saitama Medical University, 1981, Kamoda, Kawagoe City, Saitama 350-8550, Japan; and Department of Radiation Therapy, JCHO Tokyo Shinjuku Medical Center, 5-1, Tsukudocho, Shinjuku, Tokyo 162-8543, Japan. Tel: +81 492283511, Mob: +81 8019045009
nobuko_me005@yahoo.co.jp
Abstract
Background/Aim: The most severe adverse event of radiotherapy in lung cancer is radiation pneumonitis (RP). Some indices commonly used to prevent RP are evaluated based on the anatomical lung volume. The irradiation dose may be more accurately assessed by using functional lung volume. We evaluated the usefulness of computed tomography (CT) incorporating functional ventilation images acquired by the inhalation of xenon (Xe) gas (Xe-CT functional images). Patients and Methods: Two plans were created for twelve patients: volumetric modulated arc therapy (VMAT) planning using conventional chest CT images (anatomical plans) and VMAT planning using Xe-CT functional images (functional plans), and the dosimetric parameters were compared. Results: Compared to the anatomical plans, the functional plans had significantly reduced V20Gy in the high-functional lungs (p=0.005), but significant differences were not seen in the moderate-functional and low-functional lungs. Conclusion: The incorporation of Xe-CT functional images into VMAT plans enables radiotherapy planning with consideration of lung function.
Keywords: lung cancer, VMAT, radiation pneumonitis, functional radiotherapy planning, xenon
Lung cancer is the fourth most prevalent malignancy in men and the third most prevalent in women in Japan (1). Approximately 14% of patients with non-small-cell lung carcinoma (NSCLC) at the first visit have stage III disease (1). Chemoradiotherapy is standard therapy for inoperable locally advanced NSCLC of stages IIIA and IIIB; however, the therapeutic outcome of chemoradiotherapy for stage III NSCLC is poor. The five-year survival rate is 15.1% (2) and the average overall survival is 16-30 months. Although dose escalation for improving the therapeutic outcomes of radiotherapy for lung cancer has been attempted, the increase in adverse events due to high irradiation doses is a significant issue. In an attempt to reduce adverse events, the use of intensity modulated radiation therapy (IMRT) or volumetric modulated arc therapy (VMAT) has increased. The most severe adverse event of radiotherapy for lung cancer is radiation pneumonitis (RP). Following chemoradiotherapy for locally advanced lung cancer, symptomatic RP occurs in 5-50% of patients and leads to fatal pneumonia in 1-2% of patients. It has been reported that development of RP is dependent on the irradiated volume and the irradiation dose of the lungs. V20Gy (percentage of total lung volume irradiated with 20 Gy or higher), VS5Gy (absolute lung volume spared from a dose of 5 Gy or higher), and mean lung dose are used as dose evaluation indices (3-7). However, these indices are evaluated based on the anatomical lung volume by computed tomography (CT) images, not considering lung function. However, some patients with lung cancer have pulmonary diseases such as chronic obstructive pulmonary disease (COPD). Therefore, the irradiation dose may be more accurately assessed in radiotherapy planning by using functional lung volume rather than anatomical lung volume. Regarding the correlation between lung function and the occurrence of adverse events, Robnett et al. and Warner et al. reported that patients with poor pretreatment respiratory function tended to present with symptoms of pulmonary impairment after radiotherapy (8,9). Yorke et al. demonstrated that irradiation dose to the lower portion of the lungs was more important than that to the upper portion of the lungs for prediction of adverse events of the lungs after radiotherapy (10). Therefore, radiotherapy planning in consideration with lung function has been attempted in recent years. Several methods of acquiring functional ventilation images for radiotherapy planning have been reported such as the technique of using low attenuation areas (LAAs) on CT images (11,12), and the use of four-dimensional CT (4D-CT) (13-17) and 99 mTc-macroaggregated albumin (MAA) single photon emission CT (SPECT) (15,18-21). In this study, we focused on acquiring chest CT images with patients inhaling xenon (Xe). Inhalation of Xe raises the CT value and the extent of the increase reflects the distribution of Xe. Unlike the LAA method or 4D-CT, this method, established by Honda et al. (22), evaluates lung ventilation directly. Moreover, the advantage of this method over SPECT is that higher-resolution images are obtained. To date, there has been no study on the use of chest CT images acquired by Xe inhalation (Xe-CT functional images) and evaluation of lung ventilation for radiotherapy planning. In this study, we investigated a technique to create VMAT planning using Xe-CT functional images (functional plans). In addition, we compared it with VMAT planning using conventional chest CT images (anatomical plans) and evaluated the usefulness of functional radiotherapy planning using Xe inhalation.
Patients and Methods
The present study was approved by the Ethics Committee of the Saitama Medical Center, Saitama Medical University (1191-II~V). Twelve patients with lung cancer who underwent radiotherapy between April 2017 and March 2018 were enrolled in this study. We obtained written informed consent from all patients. This study was a treatment planning study and the method developed in this study was not used in actual clinical practice.
The participants inhaled Xe and Xe-CT functional images were acquired with a dual-energy (tube voltages of 80 kV and 140 kV/Sn) CT scanner, Somatom Definition Flash (Siemens Healthcare, Forchheim). The patients were placed in a supine position with the upper limbs lifted, which is the same position maintained during CT scans for radiotherapy planning. A face mask (King Systems, Nobelsville, Ind) was fixed by a strap to cover the mouth and nose of the patient and was connected to an outlet and inlet of a xenon control system Az-726 V Xetron VI (Anzai Medical). The patients were scanned in dual-energy mode during breath hold after a vital-capacity inspiration of a mixture of 35% xenon and 65% oxygen using the xenon control system. The density of Xe-54 was 5.9 kg/m3 (standard condition), approximately 4.6 times the density of air. The ventilated lung regions showed a high CT value. Since details of the method for obtaining Xe-CT functional images have been already described (22,23), only a brief explanation is given herein. Syngo Multimodality Workplace (Siemens Healthcare) software was used for the distinction of Xe on CT images. Following this, using the Velocity (Varian Medical Systems) software platform, CT value histograms of Xe in Xe-CT images were acquired. The upper limit of the small peak on the left tail of the pixel-value histograms of Xe-CT images of the whole lung was detected and it was applied as a threshold to remove the pulmonary arteries and veins from the Xe-CT images, and then Xe-CT functional images were created. We divided the area of the CT-value histograms showing functional lungs into three equal areas based on volume. The three equal areas were termed high-functional, moderate-functional, and low-functional lungs from the lung regions showing high CT values. CT images for radiotherapy planning were acquired using an Optima CT 660 Pro Advance scanner (GE Healthcare). For the fusion of Xe-CT functional images and radiotherapy planning CT images, deformable image registration (DIR) was performed using Velocity software. Xe-CT functional images acquired during maximal inhalation were fused with the radiotherapy planning CT images acquired during regular breathing (Figure 1).
Eclipse (Varian Medical Systems) software was used for VMAT planning. The gross tumor volume, clinical target volume, planning target volume (PTV), and organs at risk (OARs), including the heart, spinal cord, esophagus, and lungs, were contoured on radiotherapy planning CT images. The spinal cord and esophagus were expanded by 5 mm to generate the planning organ at risk volumes (PRVs). Anatomical plans and functional plans were created for all 12 patients. We prescribed a mean dose of 60 Gy in 30 fractions to the PTV. Table I lists the dose-volume constraints for each structure in both plans. Except for the lungs, the parameters of the anatomical plans and the functional plans were set to be the same for comparison. The difference between the two plans lied in how the dose was constrained to the lungs. In the anatomical plans, dose constraints were set uniformly over the entire lung, whereas in the functional plans, dose constraints were set in each of the three areas of high-functional, moderate-functional, and low-functional lungs. The weight of the dose constraint for V18Gy was set so that the high-functional lungs were prioritized over the moderate-functional and low-functional lungs with the greatest weight for the high-functional lungs.
The following items were evaluated: (1) among the lung domains, (a) the mean dose to the high-functional, moderate-functional, low-functional, and total-functional lungs (Gy); and (b) the V5Gy, V20Gy, V40Gy, and V50Gy of the high-functional lungs and the V20Gy of the moderate-functional and low-functional lungs (%) (VxGy: percentage of volume receiving ≥ x Gy); (2) among the PTV domains, (a) the homogeneity index (H.I.) of the PTV (H.I.=[D2%-D98%]/D50%, DX%; dose covering X% of the PTV) and (b) the conformity index (C.I.) of the PTV (C.I.=the total volume covered by the prescription isodose/PTV); (3) among the OARs, (a) the V40Gy of the heart (%), (b) the maximum dose to the spine cord (Gy), and (c) the mean dose to the esophagus (Gy); and (4) the monitor units. Dose-volume metrics were calculated using functional CT images. For analysis of the correlation between the two plans, the Wilcoxon signed-rank test was used for each index. Statistical significance was defined as p<0.05. SPSS for Windows version 23 software (IBM) was used for all statistical analyses.
Results
Patient characteristics are shown in Table II. The tumors were distributed in the upper or middle lobes, and no tumors were seen in the lower lobes. Sufficient inhalation of Xe during the acquirement of Xe-CT functional images was confirmed in all patients, and no adverse events due to Xe-CT scanning were noted. Suitable fusion of the Xe-CT functional images and radiotherapy planning CT images was performed for all patients; thus, creating both the anatomical and functional plans.
Mean dose of the high-functional, moderate-functional, low-functional, and total functional lungs in the two plans are shown in Figure 2. Compared to the anatomical plans, the functional plans had significantly reduced mean doses to the high-functional, moderate-functional, low-functional, and total functional lungs (p=0.002, p=0.003, p=0.002, and p=0.002, respectively), and the corresponding differences were 0.32, 0.27, 0.20, and 0.27 Gy, respectively.
The V5Gy, V20Gy, V40Gy, and V50Gy of the high-functional lungs and the V20Gy of the moderate-functional and low-functional lungs in the anatomical and functional plans are shown in Figure 3. Compared to the anatomical plans, the functional plans had significantly reduced V5Gy, V20Gy, V40Gy, and V50Gy of the high-functional lungs (p=0.005, p=0.005, p=0.002, and p=0.006, respectively), and the corresponding differences were 0.7%, 1.3%, 0.3%, and 0.1%, respectively. Although the V20Gy was reduced by 0.7% and 0.5% in the moderate-functional and low-functional lungs, respectively, upon comparison between the functional and anatomical plans, a significant difference was not seen.
The dosimetric parameters of PTV, OARs, and monitor units in the anatomical and functional plans are shown in Table III. H.I. was significantly increased in the functional plans compared to that in the anatomical plans (p=0.002). In contrast, no significant difference was seen in the C.I. among the two plans. In comparison to that in the anatomical plans, a significant reduction in the V40Gy of the heart was observed in the functional plans (p=0.007). Significant differences were not observed in the maximum doses to the spinal cord or the mean doses to the esophagus. Compared to the anatomical plans, the functional plans showed a significant reduction in monitor units (p=0.050).
Discussion
In Japan, Xe inhalation is only approved for use in Xe/CT cerebral-blood-flow evaluation. The maximum concentration of Xe allowed is 35%; therefore, the concentration used in this study is within the approved levels. Acquisition of Xe-CT functional images without any adverse events related to Xe inhalation was possible in our study.
In this study, the Velocity software was used for DIR of the Xe-CT functional images and radiotherapy planning CT images. Regarding the validation of DIR, we visually confirmed that the accuracy conformed to the guidelines (24,25). There are simple quantitative evaluations such as the dice coefficient and more accurate quantitative evaluations as further evaluations. However, since the tumors were distributed in the upper or middle lobes with relatively little respiratory movement in this study, we only confirmed that visually suitable fusion images were obtained that could be used for radiotherapy planning.
Compared to that in the anatomical plans, the mean dose in the functional plans was reduced from 13.04 Gy to 12.72 Gy and the V20Gy was reduced from 25.9% to 24.6% in the high-functional lungs. Although significant reductions were observed in the mean doses to the high-functional, moderate-functional, and low-functional lungs, the difference was the largest in the high-functional lungs. For IMRT and VMAT planning, Kimura et al. used the LAA method for acquiring functional images, and a radiotherapy dose of 70 Gy was prescribed; in the high-functional lungs, a mean dose reduction of 0.3 Gy was reported in the IMRT plans and of 0.5 Gy was reported in VMAT plans, and a significant reduction in V20Gy was also reported (11). Yamamoto et al. used 4D-CT for lung functional imaging and prescribed a radiotherapy dose of 74 Gy; in the high-functional lungs, the mean dose reduced from 13.4 Gy to 11.6 Gy in the IMRT plans and from 15.9 Gy to 13.9 Gy in the VMAT plans, and a reduction in V20Gy was also reported (13). Huang et al. used 4D-CT for lung functional imaging and prescribed a radiotherapy dose of 50-74 Gy; in the high-functional lungs, the mean dose reduced from 8.61 Gy to 7.38 Gy and the V20Gy reduced from 15.22% to 12.57% in the IMRT plans (14). Thus, adding dose restrictions to conventional radiotherapy planning and considering lung function by using Xe-CT functional images improved dose distribution in the lungs, especially in the high-functional lungs. In PTV analysis, although a significant increase was noted in the H.I. in the functional plans, no significant difference was seen in the C.I. In another study, Kimura et al. reported no significant changes in the H.I. or C.I. in their functional plans (11). In contrast, Yamamoto et al. reported a significant increase in the H.I. and a significant decrease in the C.I. in their functional plans (13). Compared to the anatomical plans, the functional plans showed a significantly reduced V40Gy of the heart (a difference of 0.2%), and no significant differences were seen between the two plans in the maximum dose to the spinal cord and the mean dose to the esophagus. Kimura et al. and Yamamoto et al. reported no significant changes in the average and maximum doses to OARs other than the lungs in their functional plans (11,13). We believe that our technique of radiotherapy planning considering lung function reduced the absorbed dose of the high-functional lungs and maintained the dose distribution of the PTV and OARs at a similar level with that in the anatomical plans.
The advantage of our technique is that functional images that directly evaluate the lung ventilation can be obtained as high-resolution images. The functional images by the LAA method and the method using 4D-CT are acquired by identifying functional lung regions with CT images, and their evaluation for lung function is indirect. The SPECT method can obtain functional images that directly evaluate lung function, but the resolution of the obtained functional images is low. Xe-CT functional images are acquired by actual Xe inhalation by the patients, so that lung ventilation is directly evaluated and also their resolution is high. In addition to this, our technique has another advantage of using VMAT for radiotherapy planning. VMAT is a technique which demonstrates favorable dose distribution and short treatment delivery time. Valakh et al. compared a single 360˚ coplanar arc VMAT plan with IMRT and a single 180˚ coplanar arc VMAT plan for a small lung nodule that was centrally located in close proximity to the mediastinal structures, and reported that VMAT improved esophageal sparing compared to IMRT and a single 180˚ coplanar arc VMAT plan had the highest dose conformity (26). In this study, IMRT planning was not assessed, however, better dose distribution could be obtained by VMAT planning compared to IMRT planning. On the other hand, in order to acquire Xe-CT functional images, there is a demerit that patients take CT images inhaling Xe. The LAA method and the method using 4D-CT use radiotherapy planning CT images to obtain functional images and they do not require taking additional images. Our technique has a disadvantage that additional CT scan, and the fusion of Xe-CT functional images and radiotherapy planning CT images are required.
Honda et al. reported uneven Xe distribution in Xe-CT functional images in one patient with COPD. In this patient, no abnormality was observed on conventional lung CT images acquired by breath-holding following maximal inhalation. They reported that the abnormality in ventilation could be detected on Xe-CT functional images but could not be detected using morphologic data. Further evaluation of the utility of this technique in these patients is required.
We acknowledge there are some limitations in this study. First, the number of cases is small. Second, there is a bias in the patient background. The subjects of this study had relatively good lung function enough to receive definitive radiotherapy. Coincidentally, the tumors were distributed in the upper or middle lobes with relatively little respiratory movement compare to lower lobes. Third, the functional lungs were divided into three areas based on the volume of each case, which was only classified by relative evaluation within the individual, and high-functional lungs were not defined with something like a threshold. Since the number of cases in this study was small and the purpose of this study was to develop a technique of radiotherapy planning using functional ventilation images acquired by Xe-CT, no patient selection criteria were set for lung function or tumor location, and we evaluated three areas based on the volume of each case for convenience. Even in such a patient background, it was confirmed that the high-functional lungs and the low-functional lungs could be separated into different doses. Further research is required for patients with lower lung function with COPD or lower lobe tumors, which might lead to dose difference of radiobiological significance. Moreover, by accumulating the number of cases, it might be possible to set an optimum threshold value to define high-functional lungs.
We devised a method of VMAT radiotherapy planning using functional images obtained by inhalation of Xe. This method has not been reported in any previous study. The plans that used this technique showed a reduction in absorbed doses, predominantly in the doses to the high-functional lungs. In addition, the dose distribution of the PTV or OARs were maintained at a similar level with that in the conventional anatomical plans. Thus, we believe that VMAT plans with Xe-CT functional images by the technique used in this study enable radiotherapy planning with consideration of lung function. Further research is required to identify patients who would benefit the most from this method.
Conflicts of Interest
The Authors have no financial relationships to disclose.
Authors’ Contributions
T.T. and N.H. developed the concept. N.U., T.T., S.H., M.H., M.S., S.K., R.S., K. N., and T.Y. researched the data. N.U. wrote the manuscript. T.T., S.H., M. H., and M.S. contributed to the discussion of the manuscript. N.U., T.T., S.H., M.H., M.S., S.K., R.S., K. N., T.Y., W.W., M. S. and N.H. had full access to all the data in the study and reviewed the manuscript.
References
1
. Cancer registry and statistics. Tokyo, Cancer Information Service, National Cancer Center.. Available at: https://ganjoho.jp/reg_stat/statistics/stat/summary.html.
2
Aupérin A
,
Le Péchoux C
,
Rolland E
,
Curran WJ
,
Furuse K
,
Fournel P
,
Belderbos J
,
Clamon G
,
Ulutin HC
,
Paulus R
,
Yamanaka T
,
Bozonnat MC
,
Uitterhoeve A
,
Wang X
,
Stewart L
,
Arriagada R
,
Burdett S
&
Pignon JP
. Meta-analysis of concomitant versus sequential radiochemotherapy in locally advanced non-small-cell lung cancer. J Clin Oncol.
28(13)
2181
- 2190
2010.
PMID:
20351327.
DOI:
10.1200/JCO.2009.26.2543
3
Tsujino K
,
Hirota S
,
Endo M
,
Obayashi K
,
Kotani Y
,
Satouchi M
,
Kado T
&
Takada Y
. Predictive value of dose-volume histogram parameters for predicting radiation pneumonitis after concurrent chemoradiation for lung cancer. Int J Radiat Oncol Biol Phys.
55(1)
110
- 115
2003.
PMID:
12504042.
DOI:
10.1016/s0360-3016(02)03807-5
4
Tsujino K
,
Hashimoto T
,
Shimada T
,
Yoden E
,
Fujii O
,
Ota Y
,
Satouchi M
,
Negoro S
,
Adachi S
&
Soejima T
. Combined analysis of V20, VS5, pulmonary fibrosis score on baseline computed tomography, and patient age improves prediction of severe radiation pneumonitis after concurrent chemoradiotherapy for locally advanced non-small-cell lung cancer. J Thorac Oncol.
9(7)
983
- 990
2014.
PMID:
24922010.
DOI:
10.1097/JTO.0000000000000187
5
Kimura T
,
Togami T
,
Takashima H
,
Nishiyama Y
,
Ohkawa M
&
Nagata Y
. Radiation pneumonitis in patients with lung and mediastinal tumours: a retrospective study of risk factors focused on pulmonary emphysema. Br J Radiol.
85(1010)
135
- 141
2012.
PMID:
21385918.
DOI:
10.1259/bjr/32629867
6
Marks LB
,
Bentzen SM
,
Deasy JO
,
Kong FM
,
Bradley JD
,
Vogelius IS
,
El Naqa I
,
Hubbs JL
,
Lebesque JV
,
Timmerman RD
,
Martel MK
&
Jackson A
. Radiation dose-volume effects in the lung. Int J Radiat Oncol Biol Phys.
76(3 Suppl)
S70
- S76
2010.
PMID:
20171521.
DOI:
10.1016/j.ijrobp.2009.06.091
7
Graham MV
,
Purdy JA
,
Emami B
,
Harms W
,
Bosch W
,
Lockett MA
&
Perez CA
. Clinical dose-volume histogram analysis for pneumonitis after 3D treatment for non-small cell lung cancer (NSCLC). Int J Radiat Oncol Biol Phys.
45(2)
323
- 329
1999.
PMID:
10487552.
DOI:
10.1016/s0360-3016(99)00183-2
8
Robnett TJ
,
Machtay M
,
Vines EF
,
McKenna MG
,
Algazy KM
&
McKenna WG
. Factors predicting severe radiation pneumonitis in patients receiving definitive chemoradiation for lung cancer. Int J Radiat Oncol Biol Phys.
48(1)
89
- 94
2000.
PMID:
10924976.
DOI:
10.1016/s0360-3016(00)00648-9
9
Warner A
,
Dahele M
,
Hu B
,
Palma DA
,
Senan S
,
Oberije C
,
Tsujino K
,
Moreno-Jimenez M
,
Kim TH
,
Marks LB
,
Rengan R
,
De Petris L
,
Ramella S
,
De Ruyck K
,
De Dios NR
,
Bradley JD
&
Rodrigues G
. Factors associated with early mortality in patients treated with concurrent chemoradiation therapy for locally advanced non-small cell lung cancer. Int J Radiat Oncol Biol Phys.
94(3)
612
- 620
2016.
PMID:
26867890.
DOI:
10.1016/j.ijrobp.2015.11.030
10
Yorke ED
,
Jackson A
,
Rosenzweig KE
,
Merrick SA
,
Gabrys D
,
Venkatraman ES
,
Burman CM
,
Leibel SA
&
Ling CC
. Dose-volume factors contributing to the incidence of radiation pneumonitis in non-small-cell lung cancer patients treated with three-dimensional conformal radiation therapy. Int J Radiat Oncol Biol Phys.
54(2)
329
- 339
2002.
PMID:
12243805.
DOI:
10.1016/s0360-3016(02)02929-2
11
Kimura T
,
Nishibuchi I
,
Murakami Y
,
Kenjo M
,
Kaneyasu Y
&
Nagata Y
. Functional image-guided radiotherapy planning in respiratory-gated intensity-modulated radiotherapy for lung cancer patients with chronic obstructive pulmonary disease. Int J Radiat Oncol Biol Phys.
82(4)
e663
- e670
2012.
PMID:
22245191.
DOI:
10.1016/j.ijrobp.2011.08.016
12
Defraene G
,
van Elmpt W
&
De Ruysscher D
. Regional lung avoidance by CT numbers to reduce radiation-induced lung damage risk in non-small-cell lung cancer: a simulation study. Acta Oncol.
59(2)
201
- 207
2020.
PMID:
31549562.
DOI:
10.1080/0284186X.2019.1669814
13
Yamamoto T
,
Kabus S
,
von Berg J
,
Lorenz C
&
Keall PJ
. Impact of four-dimensional computed tomography pulmonary ventilation imaging-based functional avoidance for lung cancer radiotherapy. Int J Radiat Oncol Biol Phys.
79(1)
279
- 288
2011.
PMID:
20646852.
DOI:
10.1016/j.ijrobp.2010.02.008
14
Huang TC
,
Hsiao CY
,
Chien CR
,
Liang JA
,
Shih TC
&
Zhang GG
. IMRT treatment plans and functional planning with functional lung imaging from 4D-CT for thoracic cancer patients. Radiat Oncol.
8
3
2013.
PMID:
23281734.
DOI:
10.1186/1748-717X-8-3
15
Kida S
,
Bal M
,
Kabus S
,
Negahdar M
,
Shan X
,
Loo BW Jr
,
Keall PJ
&
Yamamoto T
. CT ventilation functional image-based IMRT treatment plans are comparable to SPECT ventilation functional image-based plans. Radiother Oncol.
118(3)
521
- 527
2016.
PMID:
26922488.
DOI:
10.1016/j.radonc.2016.02.019
16
Sharifi H
,
McDonald GC
,
Lee JK
,
Ajlouni MI
,
Chetty IJ
&
Zhong H
. Four-dimensional computed tomography-based biomechanical measurements of pulmonary function and their correlation with clinical outcome for lung stereotactic body radiation therapy patients. Quant Imaging Med Surg.
9(7)
1278
- 1287
2019.
PMID:
31448213.
DOI:
10.21037/qims.2019.07.03
17
Vinogradskiy Y
,
Rusthoven CG
,
Schubert L
,
Jones B
,
Faught A
,
Castillo R
,
Castillo E
,
Gaspar LE
,
Kwak J
,
Waxweiler T
,
Dougherty M
,
Gao D
,
Stevens C
,
Miften M
,
Kavanagh B
,
Guerrero T
&
Grills I
. Interim analysis of a two-institution, prospective clinical trial of 4DCT-ventilation-based functional avoidance radiation therapy. Int J Radiat Oncol Biol Phys.
102(4)
1357
- 1365
2018.
PMID:
30353873.
DOI:
10.1016/j.ijrobp.2018.07.186
18
Seppenwoolde Y
,
Engelsman M
,
De Jaeger K
,
Muller SH
,
Baas P
,
McShan DL
,
Fraass BA
,
Kessler ML
,
Belderbos JS
,
Boersma LJ
&
Lebesque JV
. Optimizing radiation treatment plans for lung cancer using lung perfusion information. Radiother Oncol.
63(2)
165
- 177
2002.
PMID:
12063006.
DOI:
10.1016/s0167-8140(02)00075-0
19
Christian JA
,
Partridge M
,
Nioutsikou E
,
Cook G
,
McNair HA
,
Cronin B
,
Courbon F
,
Bedford JL
&
Brada M
. The incorporation of SPECT functional lung imaging into inverse radiotherapy planning for non-small cell lung cancer. Radiother Oncol.
77(3)
271
- 277
2005.
PMID:
16274762.
DOI:
10.1016/j.radonc.2005.08.008
20
Lee HJ Jr
,
Zeng J
,
Vesselle HJ
,
Patel SA
,
Rengan R
&
Bowen SR
. Correlation of functional lung heterogeneity and dosimetry to radiation pneumonitis using perfusion SPECT/CT and FDG PET/CT imaging. Int J Radiat Oncol Biol Phys.
102(4)
1255
- 1264
2018.
PMID:
30108002.
DOI:
10.1016/j.ijrobp.2018.05.051
21
Thomas HMT
,
Zeng J
,
Lee HJ Jr
,
Sasidharan BK
,
Kinahan PE
,
Miyaoka RS
,
Vesselle HJ
,
Rengan R
&
Bowen SR
. Comparison of regional lung perfusion response on longitudinal MAA SPECT/CT in lung cancer patients treated with and without functional tissue-avoidance radiation therapy. Br J Radiol.
92(1103)
20190174
2019.
PMID:
31364397.
DOI:
10.1259/bjr.20190174
22
Honda N
,
Osada H
,
Watanabe W
,
Nakayama M
,
Nishimura K
,
Krauss B
&
Otani K
. Imaging of ventilation with dual-energy CT during breath hold after single vital-capacity inspiration of stable xenon. Radiology.
262(1)
262
- 268
2012.
PMID:
22025733.
DOI:
10.1148/radiol.11110569
23
Yanagita H
,
Honda N
,
Nakayama M
,
Watanabe W
,
Shimizu Y
,
Osada H
,
Nakada K
,
Okada T
,
Ohno H
,
Takahashi T
&
Otani K
. Prediction of postoperative pulmonary function: preliminary comparison of single-breath dual-energy xenon CT with three conventional methods. Jpn J Radiol.
31(6)
377
- 385
2013.
PMID:
23547011.
DOI:
10.1007/s11604-013-0202-z
24
Japanese Society for Radiation Oncology QA Committee DIR Guideline Working Group
. DIR guideline 2018. Tokyo, Kito S (eds.).. Available at: https://www.jastro.or.jp/medicalpersonnel/guideline/dir_v3.pdf.
25
Brock KK
,
Mutic S
,
McNutt TR
,
Li H
&
Kessler ML
. Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132. Med Phys.
44(7)
e43
- e76
2017.
PMID:
28376237.
DOI:
10.1002/mp.12256
26
Valakh V
,
Chan P
,
D’Adamo K
&
Micaily B
. Early-stage central lung cancer and volumetric modulated arc therapy: a dosimetric case study with literature review. Anticancer Res.
33(10)
4491
- 4495
2013.
PMID:
24123020.